AEOLUS-PRO-100 Dental Electric Aspiration System
AEOLUS-PRO-100 Dental Electric Aspiration System:Dental Negative Pressure Suction Machine is specially designed for professional treatment scenarios in dental clinics and hospitals. Serving as an essential supporting device for dental treatment units during various dental procedures such as tooth drilling, this high-performance oral suction machine effectively aspirates saliva, blood, oral debris, and aerosol mist generated by high-speed turbine handpieces. Integrated with three-in-one functions
Interview
Capacity:
Suitable for use with 90–100 dental chair units.
Equipment Purpose:
Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g., tooth drilling) in hospitals or dental clinics; also features a water vapor separation and automatic discharge function.
Operating Conditions:
Environmental Conditions:
Technical Specifications:
| Dental Air-Ring Suction Pump — Technical Specification | |||
| Section | Item / Parameter | Specification | Remarks |
| 1. General / Capacity | |||
| 1. General / Capacity | Suitable Capacity | Suitable for use with 90–100 dental chair units. | |
| 2. Equipment Purpose | |||
| 2. Equipment Purpose | Main Function | Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g. tooth drilling) in hospitals or dental clinics. | |
| 2. Equipment Purpose | Additional Function | Also features a water vapor separation and automatic discharge function. | |
| 3. Operating Conditions - Environmental | |||
| 3. Operating Conditions - Environmental | Ambient Temperature | 5 °C – 40 °C | |
| 3. Operating Conditions - Environmental | Relative Humidity | ≤ 70 % | |
| 3. Operating Conditions - Environmental | Atmospheric Pressure | 86 – 106 kPa | |
| 3. Operating Conditions - Environmental | Ventilation Flow Rate | ≥ 0.50 m³/s | |
| 3. Operating Conditions - Environmental | Machine Room Area | ≥ 20 m² | |
| 3. Operating Conditions - Power Supply | |||
| 3. Operating Conditions - Power Supply | Rated Voltage | AC 380 V ±10 % | |
| 3. Operating Conditions - Power Supply | Frequency | 50 Hz | |
| 3. Operating Conditions - Power Supply | Rated Power | ≤ 25 kW | |
| 4. Technical Specifications | |||
| 4. Technical Specifications | Suction Pump Type | Air-ring suction pump; motor speed: ≥ 2800 r/min; capable of instantaneously generating a high suction flow rate compliant with dental application standards without any waiting time. | |
| 4. Technical Specifications | Maximum Vacuum | –22 kPa | |
| 4. Technical Specifications | Suction Flow Rate | When operating at a negative pressure of –140 mBar, the suction capacity per unit pump is ≤ 5500 L/min; the maximum total suction capacity of the unit is ≤ 22,000 L/min; supports up to 100 dental chair stations. | |
| 4. Technical Specifications | Suction System | Features an integrated modular semi-dry suction system with independent wastewater pumps capable of continuous, active wastewater discharge. | |
| 4. Technical Specifications | Control System | The Siemens PLC intelligent control system, when combined with an intelligent management system, enables independent control of the operating status of each vacuum pump; this allows each vacuum pump to be started and stopped individually and operated in a rotating sequence, ensuring each pump operates for the same duration and thereby extending its service life. | |
| 4. Technical Specifications | Protection Functions | The control system incorporates multiple circuit protection functions, including an automatic delayed shutdown feature. | |
| 4. Technical Specifications | Negative Pressure Balancing | Built-in mechanical negative pressure balancing valve maintains the negative pressure within the optimal range of 140–180 mBar, preventing damage to the patient's oral mucosa caused by excessively high negative pressure. | |
| 4. Technical Specifications | Cooling | The mechanical negative pressure balancing valve ensures adequate fresh air supply for internal cooling and heat dissipation within the suction pump. | |
| 4. Technical Specifications | Noise Level | ≤ 80 dB | |
| 4. Technical Specifications | External Dimensions | ≤ 140 × 80 × 120 cm (main unit); ≤ 60 × 60 × 112 cm (separate tank) | |
| 4. Technical Specifications | Net Weight | ≤ 320 kg | |
![]()

Copy product links
Long by picture save/share


0757-88389552
Monday to Saturday
cherish@antar.cn