DP-1 Periodontal pocket depth detector
DP-1 Periodontal pocket depth detector:The Periodontal Pocket Depth Meter offers high accuracy and excellent repeatability in measuring periodontal pocket depth; it can collect patient-specific periodontal pocket data and transmit it to its companion software, which processes the data in real time, outputs the examination results, and records them. The examination results are not influenced by the visual perception of healthcare professionals; furthermore, the external hardware device is equippe
Interview
Product Technical Specifications
Probe length :≥15 20°C,40% humidity (mm)
Angle of the probe’s working end: 60° +5 20°C,40% humidity (mm)
Minimum effective detection depth: 14 20°C,40% humidity (mm)
Error in detection depth: <0.3 20°C,40% humidity (mm)
Repeatability error in detection depth: <0.3 20°C,40% humidity (mm)
Operating temperature:5-40℃ Both low and high temperatures can affect a patient's comfort.
a) Classification by type of electric shock protection: Class II;
b) Classification by degree of electric shock protection: includes an application part of Type B;
c) Classification by level of protection against liquid ingress: general-purpose equipment; foot switch: IPX1;
d) Classification by safety level when used in environments containing flammable anesthetic gases mixed with air, or flammable anesthetic gases mixed with oxygen or nitrous oxide: equipment not suitable for use under these conditions;
e) Classification by operating mode: continuous operation;
f) Equipment rated voltage and frequency: Safe Extra Low Voltage (SELV), DC 5 V
The system consists of a set of probe devices, a set of foot pedals, and embedded software.
Software Introduction
Computer Hardware Requirements:
Minimum system requirements:
4 GB RAM;512 MB free disk space;one available USB 2.0 high-speed interface,;CPU clock speed> 2 GHz; monitor resolution ≥ 1280 × 960;
Recommended system requirements:
8 GB RAM; 512 MB free disk space; one SSD solid-state drive; one available USB 2.0 high-speed interface; CPU clock speed> 2 × 2 GHz,;monitor resolution 1920 × 1280.
Software Requirements: Windows 10;.NET Framework 4.8 or later.
Please obtain and install this software from the manufacturer. If the software is corrupted or unresponsive, please contact the manufacturer's customer support for assistance.
Q: What are the probe specifications?
A: Probe length ≥ 15 mm, working end angle 60° + 5, minimum effective detection depth 14 mm.
Q: What is the detection accuracy?
A: Error in detection depth < 0.3 mm; repeatability error < 0.3 mm.
Q: What are the operating temperature and power requirements?
A: Operating temperature 5–40°C. Rated voltage: Safe Extra Low Voltage (SELV), DC 5 V.
Q: What are the safety classifications?
A: Class II electric shock protection, Type B application part, general-purpose equipment with foot switch IPX1, continuous operation. Not suitable for flammable anesthetic gas environments.
Q: What is included in the system?
A: A set of probe devices, a set of foot pedals, and embedded software.
Q: What are the software and computer requirements?
A: Windows 10 with .NET Framework 4.8 or later. Minimum: 4 GB RAM, 512 MB free disk space, USB 2.0, CPU > 2 GHz, resolution ≥ 1280×960. Recommended: 8 GB RAM, SSD, CPU > 2×2 GHz, resolution 1920×1280.
| Item | Specification | Test Condition |
|---|---|---|
| Probe Length | ≥ 15 mm | 20°C, 40% humidity |
| Angle of the Probe's Working End | 60° + 5 | 20°C, 40% humidity |
| Minimum Effective Detection Depth | 14 mm | 20°C, 40% humidity |
| Error in Detection Depth | < 0.3 mm | 20°C, 40% humidity |
| Repeatability Error in Detection Depth | < 0.3 mm | 20°C, 40% humidity |
| Operating Temperature | 5–40°C | Both low and high temperatures can affect a patient's comfort |
| Item | Classification |
|---|---|
| a) By Type of Electric Shock Protection | Class II |
| b) By Degree of Electric Shock Protection | Includes an application part of Type B |
| c) By Level of Protection Against Liquid Ingress | General-purpose equipment; foot switch: IPX1 |
| d) By Safety Level in Flammable Anesthetic Gas Environments | Equipment not suitable for use in environments containing flammable anesthetic gases mixed with air, or flammable anesthetic gases mixed with oxygen or nitrous oxide |
| e) By Operating Mode | Continuous operation |
| f) Equipment Rated Voltage and Frequency | Safe Extra Low Voltage (SELV), DC 5 V |
| Item | Description |
|---|---|
| System Components | The system consists of a set of probe devices, a set of foot pedals, and embedded software |
| Item | Requirement |
|---|---|
| Minimum System Requirements | 4 GB RAM; 512 MB free disk space; one available USB 2.0 high-speed interface; CPU clock speed > 2 GHz; monitor resolution ≥ 1280 × 960 |
| Recommended System Requirements | 8 GB RAM; 512 MB free disk space; one SSD solid-state drive; one available USB 2.0 high-speed interface; CPU clock speed > 2 × 2 GHz; monitor resolution 1920 × 1280 |
| Software Requirements | Windows 10; .NET Framework 4.8 or later |
| Software Installation | Please obtain and install this software from the manufacturer. If the software is corrupted or unresponsive, please contact the manufacturer's customer support for assistance |

Copy product links