AEOLUS-pro-30 Dental electric suction machine
Dental Negative Pressure Suction Machine is specially designed for professional treatment scenarios in dental clinics and hospitals. Serving as an essential supporting device for dental treatment units during various dental procedures such as tooth drilling, this high-performance oral suction machine effectively aspirates saliva, blood, oral debris, and aerosol mist generated by high-speed turbine handpieces. Integrated with three-in-one functions of oral cleaning, mouth rinsing and sputum irrig
Interview
Capacity:
Suitable for use with 25–30 dental combined treatment units.
Equipment Purpose:
Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g., tooth drilling) in hospital or dental clinic settings; also features a water vapor separation and automatic discharge function.
Operating Conditions :
Environmental Conditions :
Technical Specifications:
Configuration List:
Negative Pressure System Main Unit ×1;
Separation Tank ×1;
Throat Clamp ×3;
Drainage Hose ×1;
Connection Adapter (Set) ×1
Q: How many dental units can it support?
A: 25–30 dental combined treatment units.
Q: What is the main purpose?
A: Aspirating saliva, blood, small particles, or other fluids during dental treatment; also separates water vapor and discharges automatically.
Q: What are the power and environment requirements?
A: AC 380 V ±10%, 50 Hz, ≥ 7 kW. Ambient 5–40°C, humidity ≤ 70%, pressure 86–106 kPa, ventilation ≥ 0.40 m³/s.
Q: What is the suction performance?
A: Motor ≥ 2800 rpm, max vacuum –22 kPa. At 0 kPa: single pump ≥ 4500 L/min, unit ≥ 13,500 L/min. At –15 kPa: single pump ≥ 2100 L/min, unit ≥ 6300 L/min. Three pumps in parallel.
Q: How is it controlled?
A: Microcomputer control system with independent pump control, rotating sequence operation, multiple circuit protection, delayed shutdown, and remote monitoring.
Q: What is included, and what are the noise, size, and weight?
A: Includes main unit, separation tank, throat clamp ×3, drainage hose, connection adapter set. Noise ≤ 79 dB(A). Size ≤ 150×65×100 cm or 50×50×140 cm. Weight ≤ 320 kg.
| Category | Item | Content |
|---|---|---|
| Capacity | Suitable for use with | 25–30 dental combined treatment units |
| Equipment Purpose | Purpose | Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g., tooth drilling) in hospital or dental clinic settings; also features a water vapor separation and automatic discharge function. |
| Environmental Conditions | Ambient Temperature | 5°C – 40°C |
| Relative Humidity | ≤ 70% | |
| Atmospheric Pressure | 86 – 106 kPa | |
| Ventilation Flow Rate | ≥ 0.40 m³/s | |
| Power Supply Requirements | Rated Voltage | AC 380 V ± 10% |
| Frequency | 50 Hz | |
| Rated Power | ≥ 7 kW | |
| Power Supply Summary | Rated Voltage: AC 380 V ± 10%; Frequency: 50 Hz; Rated Power: ≥ 7 kW | |
| Technical Specifications | Pump Type | Air-ring suction pump |
| Motor Speed | ≥ 2800 rpm | |
| Feature | Capable of instantly generating a high suction flow rate compliant with dental application standards without any waiting time | |
| Maximum Vacuum | –22 kPa | |
| High Suction Flow Rate (P = 0 kPa) | Suction capacity of a single vacuum pump ≥ 4500 L/min; total suction capacity of the unit ≥ 13,500 L/min | |
| High Suction Flow Rate (P = –15 kPa) | Suction capacity of a single vacuum pump ≥ 2100 L/min; total suction capacity of the unit ≥ 6300 L/min | |
| Parallel Operation | Three vacuum pumps are operated in parallel to accommodate up to 25–30 dental comprehensive treatment units | |
| System Features | Features an integrated modular dry suction system with an independent wastewater pump for continuous, active wastewater discharge; includes a high-efficiency stainless steel separation tank, a high-precision liquid level sensor, and an automatic flushing and cleaning system within the tank | |
| Control System | The microcomputer-controlled system, integrated with an intelligent management system, enables independent control of the operating status of each vacuum pump; each pump can be started and stopped individually or operated in a rotating sequence to ensure equal operating time for all pumps, thereby extending their service life; the system features multiple circuit protection functions and an automatic delayed shutdown function | |
| Remote Monitoring System | The remote monitoring system ensures the reliable and efficient operation of the equipment | |
| Technical Specifications | Mechanical Negative Pressure Balancing Valve | The built-in mechanical negative pressure balancing valve maintains the negative pressure within the optimal range of 140–180 mbar, preventing damage to the patient's oral mucosa caused by excessively high negative pressure; also ensures adequate fresh air supply for internal cooling and heat dissipation within the suction pump |
| Noise Level | ≤ 79 dB(A) | |
| External Dimensions | (Height × Width × Length) ≤ 150 × 65 × 100 cm (main unit) or 50 × 50 × 140 cm (separate tank) | |
| Net Weight | ≤ 320 kg |
| Item | Quantity |
|---|---|
| Negative Pressure System Main Unit | ×1 |
| Separation Tank | ×1 |
| Throat Clamp | ×3 |
| Drainage Hose | ×1 |
| Connection Adapter (Set) | ×1 |