AEOLUS-1500X Dental Electric Aspirator:Dental Negative Pressure Suction Machine is specially designed for professional treatment scenarios in dental clinics and hospitals. Serving as an essential supporting device for dental treatment units during various dental procedures such as tooth drilling, this high-performance oral suction machine effectively aspirates saliva, blood, oral debris, and aerosol mist generated by high-speed turbine handpieces. Integrated with three-in-one functions of oral c
Interview
Quantity:
Suitable for use with 4–6 dental combined treatment units.
Equipment Purpose:
Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g., tooth drilling) in hospital or dental clinic settings; also features a water vapor separation and automatic discharge function.
Operating Conditions :
1 .Environmental Conditions
Power supply requirements:
Technical Specifications:
Equipment List: :
Vacuum Pump ×1;
Exhaust Filter ×1;
Vacuum Hose ×1; Drain Hose ×1;
Exhaust Hose ×1
Q: How many dental units can it support?
A: 4–6 dental combined treatment units.
Q: What is the main purpose?
A: Aspirating saliva, blood, small particles, or other fluids during dental treatment; also separates water vapor and discharges automatically.
Q: What are the power and environment requirements?
A: AC 220 V ±10%, 50 Hz, 1.9 kW. Ambient 5–40°C, humidity ≤ 70%, pressure 86–106 kPa, ventilation ≥ 0.10 m³/s.
Q: What is the suction performance?
A: Motor ≥ 2800 rpm, max vacuum –19 kPa. At –15 kPa, suction flow rate ≥ 1000 L/min, supporting up to 6 dental units.
Q: How is it controlled and protected?
A: Microcomputer controller with multiple circuit protection, delayed shutdown, built-in negative pressure balancing valve, wastewater pump, drainage diaphragm valve, and silver-ion exhaust filter.
Q: What are the noise, size, and weight?
A: Noise ≤ 68 dB(A). Size ≤ 92×44×70 cm. Weight ≤ 70 kg.
| Category | Item | Content |
|---|---|---|
| Quantity | Suitable for use with | 4–6 dental combined treatment units |
| Equipment Purpose | Purpose | Used as a suction power source for aspirating saliva, blood, small particles, or other fluids during dental treatment procedures (e.g., tooth drilling) in hospital or dental clinic settings; also features a water vapor separation and automatic discharge function. |
| Environmental Conditions | Ambient Temperature | 5°C – 40°C |
| Relative Humidity | ≤ 70% | |
| Atmospheric Pressure | 86 – 106 kPa | |
| Ventilation Flow Rate | ≥ 0.10 m³/s | |
| Power Supply Requirements | Rated Voltage | AC 220 V ± 10% |
| Frequency | 50 Hz | |
| Rated Power | 1.9 kW | |
| Power Supply Summary | Rated Voltage: AC 220 V ± 10%; Frequency: 50 Hz; Rated Power: 1.9 kW | |
| Technical Specifications | Pump Type | Air-ring suction pump |
| Motor Speed | ≥ 2800 rpm | |
| Feature | Capable of instantly generating a high suction flow rate compliant with dental application standards without any waiting time | |
| Maximum Vacuum | –19 kPa | |
| High Suction Flow Rate (at –15 kPa) | ≥ 1000 L/min; supports up to 6 dental combined treatment units | |
| System Features | Features an integrated modular wet-suction structure equipped with a two-stage water–gas separation system and an active gas–liquid separator | |
| Gas–Liquid Separator | Incorporates a two-stage separation function; liquid handling capacity ≥ 20 L/min | |
| Casing & Components | The entire machine casing and components are manufactured from electroplated anti-corrosive aluminum alloy or high-strength anti-corrosive plastic, extending the service life of the equipment | |
| Control System | Microcomputer controller with multiple circuit protection functions and an automatic delayed shutdown feature | |
| Mechanical Negative Pressure Balancing Valve | Built-in mechanical negative pressure balancing valve ensures the negative pressure remains within the optimal range of –14 kPa to –18 kPa, preventing excessive negative pressure from damaging the patient's oral mucosa; also ensures adequate fresh air supply for internal cooling and heat dissipation of the suction pump, while preventing damage to the suction pump caused by pipeline orifices | |
| Wastewater Pump | Built-in wastewater pump enables continuous automatic discharge of wastewater | |
| Drainage Diaphragm Valve | Built-in drainage diaphragm valve prevents backflow of wastewater into the suction pump | |
| Exhaust Filter | Equipped with an exhaust filter featuring a built-in silver-ion-coated filter element, providing both filtration and sterilization functions | |
| Technical Specifications | Noise Level | ≤ 68 dB(A) |
| External Dimensions | (Height × Width × Depth) ≤ 92 × 44 × 70 cm | |
| Weight | ≤ 70 kg |
| Equipment | Quantity |
|---|---|
| Vacuum Pump | ×1 |
| Exhaust Filter | ×1 |
| Vacuum Hose | ×1 |
| Drain Hose | ×1 |
| Exhaust Hose | ×1 |